101-1GM100-010
4. Time format 12/24H selection
With the time format blinking, press the main button to
adjust it. Then hold the same button for 2 seconds to
confirm it and move to hours setting.
5. Hour setting
With the hour blinking, press the main button until the current hour
appears. Then hold the same button for 2 seconds to confirm it and
move to minute setting.
6. Minute setting
With the minute blinking, press the main button to
adjust it. Then hold the same button for 2 seconds to
confirm. It will shift to next digit for setting. Repeat the
above action until the minute setting is completed.
Then it will move to average-day setting.
7. Average-day setting
With the number " 1, 7, 14, 30 " blinking, press the
main button for the days of the average you would like
to read. Then hold the same button for 2 seconds to
confirm it and move to measurement unit setting.
1. Year setting
With the year format blinking, press the main button to adjust it. Then
hold the same button for 2 seconds to confirm. It will shift to next digit
for setting. Repeat the above action until the year setting is completed.
Then it will move to month setting.
2. Month setting
With the month blinking, press the main button until the current month
appears. Then hold the same button for 2 seconds to confirm it and
move to day setting.
3. Day setting
With the day blinking, press the main button to adjust it. Then hold the
same button for 2 seconds to confirm. It will shift to next digit for
setting. Repeat the above action until the day setting is completed.
Then it will move to time format setting.
PRECAUTION
- Re-cap the test strip vial immediately after removing a test strip.
- Do not reuse test strips.
- Do not use expired test strips.
- Record the date when opening a new vial of test strips. Discard the vial 3 months
after the date of opening.
- Store the test strips in a cool and dry location. Do not expose to direct sunlight or
heat.
- For detailed information, please refer to the GS100 Test Strip Package
Insert.
Rightest™
Intended Use
Blood Glucose Monitoring System
Instructions for Use
The Blood Glucose Monitoring System is intended for in vitro diagnostic use
(outside the body) for the quantitative measurement of glucose (sugar) in fresh capillary
whole blood from the fingertip, palm or forearm. It is indicated for self-testing by persons
with diabetes in the home or by healthcare professionals in healthcare facilities. Coding is
not required with the GM100 meter. The test provides plasma equivalent results. It is not
intended for the diagnosis of or screening for diabetes mellitus. It is not intended for use on
neonate blood samples. Alternate site testing (palm or forearm) in this system can be used
only during steady-state blood glucose conditions.
You may consult your professional for instructions on how to use the system
correctly. Our customer support staff is available to help assist you as well.
Rightest™
healthcare
1.
Materials Included2.
Preface
Thank you for purchasing the Rightest™ Blood Glucose Monitoring System. Please read
this manual thoroughly before you begin testing. This manual provides all the information
you need to perform the test correctly. It is very important that you test your blood glucose
levels routinely in order to avoid any health complications. The Rightest™ Blood Glucose
Monitoring System has been thoughtfully designed to provide you with the most accurate
and simplified testing system in order to make managing your diabetes easier than ever.
The Rightest™ Blood Glucose Monitoring System is manufactured by Bionime Corporation
and is supported by its authorized representatives. If you have any questions, please
contact your authorized distributor, in the Canada please contact Customer Service at
1 (800) 964-8434 (Monday through Friday 8:00 AM to 5:00 PM PST).
321
The Meter3.
Precautions4.
- Please perform a quality control test regularly to make sure the test results are accurate.
- The Rightest™ GM100 Blood Glucose Meter can only be used with the Rightest™
GS100 Blood Glucose Test strips. Strips from other manufacturers or brands will not work
with your meter and should not be used.
- The Rightest™ Blood Glucose Monitoring System is intended for self-testing. It should
not be used to diagnose diabetes mellitus.
- Rightest™ Blood Glucose Monitoring System has not been validated for use on
neonates. Therefore, it's not intended for use on neonates.
- This meter can be only tested with capillary whole blood. Do not use arterial blood.
- Wait 30 minutes before performing a test if you have moved the meter to a location with a
significant change in temperature.
- Dispose of used batteries properly.
- Please note the meter kit contains small parts like test strips which could result in a
choking hazard for children.
- Prevent water from entering the meter. Never immerse the meter or hold it under running
water.
- The minimum blood sample size for testing is 1.4µL :( )
Sample Size Example
1.0 1.4 2.0 3.0µL µL µL µL 4 µL .0
The GS100 Test Strip5.
Sample Entry
µL
Apply a drop of blood or control solution here.
The test request is only 1.4 of blood.
Electrode Contacts
Sensing signal output terminals.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol up
and toward meter.
Battery Installation6.
Test Strip Handle
Grip here to insert
test strip into meter.
View Window
This window is yellow before applying
blood sample. It gradually turns red when
filled with blood.
Please take a minimum of 1.4µL to run the test on your meter.
Blood sample size above 4.0µL might contaminate the test strip port.
Blood sample size below 1.4µL might cause inaccurate test result or might not start on
meter measurement. In this case, repeat the test with a new test strip.
Rightest™ GM100 is designed for use with GS100 strip only. Please note that the use of
other test strips might cause unexpected damage or produce inaccurate test results.
Display Window
Shows test results,
messages and related
information.
Test Strip Port
Insert test strip here until you
hear a click.
Main Button
Press to turn on/off. Switch
function screens or enter
setting mode. (There will be
a power-off period during
this process.)
Battery Cover
Slide cover to
change battery.
Meter Serial
Number
Model Number
Plastic Tape
Pull it off to activate meter.
Test Strips
200 7-
0
6
1. Meter
2. Test Strips (0/10/25 pcs)
3. Control Solutions
4. Control Solutions
5. Instructions for Use
6. Blood Glucose Test Strip
Package Insert
Rightest™
Rightest™
Rightest™ (Normal
Level)
Rightest™ (High
Level)
Rightest™
Rightest™
7. Lancing Device
8.
9. Disposable Sterile Lancets (10pcs)
10. Instruction for the lancing device
11. Log Book
12. Warranty Card + Emergency
13. CR2032 Battery (Inserted into the
meter)
14. Carrying Case
Rightest™
Clear Cap
8
8
7
5
6
Setting Up Your Meter
7.
PRECAUTION
- Danger of explosion exists if the battery is installed incorrectly .
- Please follow your local and state guidelings when disposing of your used
batteries.
Your meter comes with one CR2032, 3volt, battery installed. One new
battery will provide power to perform about 1000 tests under normal use.
Before using it, please pull off the plastic tape and press the main button
to activate the meter. The low battery symbol " " will keep appearing
on your display if your battery runs low. Please keep spares on hand and
replace the battery soon when a battery symbol appears.
You can enter setting mode using either of the following options:
1. Reload battery
After removing the battery, please press the button several times until the screen goes
blank, then follow the battery installation steps to load battery. The meter will do self test.
Press the main button to close the test and enter Setting Mode.
2. With Battery inserted
Press the main button first to turn on the meter. Then press and hold the main button for
5 seconds (the meter will be turned off during this period, please keep pressing) until
you hear a beep, indicating you have successfully entered Setting Mode. The display
screen will show setting data.
NOTE
- When initially entering the setting mode, if you press and hold the main button for
2 seconds, the screen display will turn off. Don't stop pressing the button. Keep
pressing the main button until you reenter the setting mode.
- Pressing and immediately releasing the main button allows you to change
individual settings. Pressing and holding the main button for 2 seconds while in
the setting mode confirms your settings.
- Holding the main button while in the setting mode for more than 5 seconds will
exit settings.
Turning on /off the Meter
1. Auto Power off
The meter will automatically shut-off after 2 minutes if it is not in use.
2. Manual Power off
The meter will shut-off if you press and hold the main button for a couple seconds.
3. How to turn on your Meter
1) Press the main button.
2) Insert a test strip.
PRECAUTION
" "The unit of measurement should be fixed to mmol/L .
1. Turn the meter over. Press and push battery cover to
open.
2. Install the battery. Be sure to put battery in correct
direction.
3. Slide the battery cover back until it snaps into place.
4. The meter performs a self-test and all symbols on the display will blink.
5. Press any button to exit the self-test and enter Setting Mode.
6. Set the time and date when the batteries are replaced. See Setting the
Date, Time and Unit section. Prior test results are still stored in the
memory.
8. Ending setting
After confirm the measurement unit, you'll hear a sound of " beep ". All the settings are
saved and completed and will return to time screen.
NOTE
- The meter will automatically shut-off after 2 minutes if it is not in use.
- If you would like to exit the setting mode, press and hold the main button for more
than 5 seconds. The current settings will be saved.
PRECAUTION
- For more information on how to install the clear adjustable cap, please read the
instruction manual for the lancing device.
- Use caution when loading the lancet.
- Do not touch your blood drop to the sample entry on the strip until you see the
" " appear. The meter is performing an internal test and will display " " and
" " if you apply blood too soon.
- Do not perform the blood glucose test at temperatures below 10°C (50°F) or
above 40°C (104°F), below 10% or above 90% relative humidity.
1. Pull off the depth adjustable cap.
2. Insert a new disposable lancet firmly into lancet carrier.
3. Twist off and set aside the protective cover of the disposable lancet.
4. Replace the depth adjustable cap.
5. Choose a depth of penetration by rotating the top portion of the depth adjustable cap
until the setting depth matches the window. Settings are based on skin type " " for
soft or thin skin; " " for average skin; " " for thick or calloused skin.
Alternative site testing-palm or forearm blood sampling
10. Add the blood sample when you observe the
flashing blood sample indicator.
11. Place the lancing device against the pad of your
fingertip and press the release button. The best
puncture sites are on the middle of fingers. Press the
release button.
12. Touch and hold the drop to the edge of sample entry until you hear a " beep " and the
view window is totally filled with blood. If the view window is not totally filled with blood
or the test does not start. Please discard the test strip and repeat the test with a new
test strip.
13. The meter will commence counting down. After 8 seconds, the test result will appear.
1110
13
1 2 3 4
5
6
7
14. Pull off the depth adjustable cap. Without touching the used disposable lancet, stick
the lancet tip into the protective cover.
15. Hold the release button in one hand and pull on the plunger in the other hand. This
will safely eject the used disposable lancet.
16. Discard the used disposable lancet into an appropriate puncture-proof or biohazard
container.
17. Replace the depth adjustable cap after finishing the test.
16
14
1
2
15
17
- Before using the Rightest™ System to test your blood glucose, please read all the
instructions and practice performing a test including a quality control test.
The blood glucose monitoring system gives you an opportunity to obtain blood
samples from fingertip, palm or forearm.
Rightest™
1. Using the clear cap, follow the above listed steps 1
through 4.
2. Massage the puncture area of palm or forearm for a
few seconds.
3.
clear cap
4. Then press the release button.
5. Continue holding the lancing device against palm or
forearm and gradually increase pressure for a few
seconds until the blood sample size is sufficient.
Immediately after massaging the puncture area, press
and hold the lancing device with the against
palm or forearm.
8. Take one strip from the vial. Re-cap the vial
immediately.
9. Insert the strip into the strip port on meter with the
indication symbol facing up.
6. Hold the hub in one hand and pull on the plunger in
the other hand. The device is now loaded. Release
the plunger, it will automatically move back to its
original position near the hub.
7. Wash your hands with warm soapy water and dry
thoroughly.
Performing a Blood Test
8.
PRECAUTION
- Blood samples obtained from alternate sites may yield varied results when your
blood glucose level is rapidly changing. For example, imediately after having a
meal, drink, a dose of insulin or after excercising. In these cases, only fingertip
blood should be used.
200 7-
0
6
4
Normal Level High Level
9
8
12
Pressing for
a few seconds
Releasing
button