About Quality Control Test
Performing a Quality Control Test
PRECAUTION
- When you open a new bottle of Rightest Solution, please write the discard date on the label.
Rightest Solution is good for 3 months after opening the bottle, or until the expiration date printed on
the label, whichever comes first.
®
®
Control
Control
- Replace the vial cap of control solution and close tightly right away after using of control solution sample.
- Check the expiration date before you use the control solution. ( Refer to Control Solution Package Insert. )
Example
Starting with Check Key
1. Insert the Rightest Check Key with " Check Key " word, facing up, into test strip port. The meter
quickly goes through a series of internal checks.
2. The check key allows you to make sure your meter is functioning well. You can see " " symbol
and " " symbol blinking on the screen.
The check key also allows your meter proceeding quality control test with control solution.
®
NOTE
The check key is only for functional testing of meter. If you want to make sure strip and meter
work properly, please use control solution for the test ( see page 52 for detail. )
Exp. Date 2014-06
3. If both " " symbol and " " symbol blinking appear, remove check key and insert it again.
If error screen appears again, your meter is not working correctly. Please contact Bionime authorized
distributor or Customer Service Center.